Contact Reflection
The NIRS Analyzer Pro is a single point process analysis system based on high resolution diode array technology.
It provides non-destructive analysis of solids, granules, or powders through a glass window directly in the process line without bypass – a true inline system.
Do NOT CHANGE OR REMOVE THIS ELEMENT
Contact Reflection
Specifications
Analysis time | 5 – 50 ms / integration time; average time per result: 1 – 15 seconds. |
---|---|
Wavelength range | 1100 – 1650 nm |
Detector | InGaAs diode spectrum |
Spectral dispersion | 1.1 nm / pixel |
General |
|
Lifetime of the light source | Dual lamp system MTBF = 17,500 h |
Software package (sold separately) | Vision for instrument operation and method development |
Wavelength accuracy | 0.5 nm |
Wavelength precision | < 0.02 nm |
Wavelength stability | < 0.01 nm/°C |
Noise | < 60 Micro-AU |
Random vibrations | 0.4 Grms at 10 – 150 Hz in accordance with IEC 60068-2-64 / 0.4 Grms at 10 – 1250 Hz in accordance with international Metrohm standard (additional information available on request) |
Temperature | -5 – 40 °C (23 – 104 °F). With purge -5 – 65 °C (23 – 149 °F). |
Purge | Flow rate min. 5 l/min, > 99.99% anhydrous, > 99.9% without oil and fine particles up to 0.3 μm |
Humidity | 10 – 90% relative |
Dimensions (W × H × D) | 42 × 42 × 13.5 cm (16.5 × 16.5 × 5.3 inches) + holders for the device |
Weight | 25 kg / 55 lbs |
Cabinet | 1.5 mm (lid 2.5 mm) stainless steel EN 1.4301 (SS2333) |
Protection | IP69K 1) in accordance with IEC 60529 and DIN 40050 Part 9, NT ELEC 023 |
Power supply | disconnected from network or network-dependent (recommended) / 100 – 240 VAC, 50 – 60 Hz, 2.0 A, 150 W |
1) IP6x is the highest protection against dust entering the device. IPx9K means protected against the effect of high-pressure water and/or steam cleaning at high temperature.